14.2 External I/O
392
2
In the Ext Term (external terminal) Settings screen, make settings for the
[START /EXT.IN1]
,
[STOP /EXT.IN2]
,
[PRINT /EXT.IN3]
external terminals.
3
Short circuit the terminal and GND, or input a HIGH level (3.0 to 5.0 V) or LOW
level (0 to 0.8 V) pulse wave or rectangular wave to the terminal.
Control with the LOW level of the input waveform.
Select the operation performed by the in-
strument in response to external signal in-
put.
START
Start measurement.
STOP
Stop measurement.
START/STOP
Start measurement on LOW level, and stop measurement
on HIGH level.
Print to the destination specified as the PRINT key output
destination.
SAVE
Save to the media specified for the SAVE key, according to
the specified conditions.
Default setting: START/EXT.IN1: START
STOP/EXT.IN2: STOP
PRINT/EXT.IN3: PRINT
(To open the screen: Press the
DISP
key
Press the
F7
[System]
key
Select
with the
SUB MENU
keys
Ext Term Settings Screen)