4.2 Setting Measurement Configuration (Status Settings Screen)
96
5
6
2
3
Setting Timebase 1 and 2: When using the Model 8958 16-
Ch Scanner Unit together with other input modules
To open the screen: Press the
SET
key
Select
with the
SUB MENU
keys
Status Settings screen
See
Screen Layout(p. 28), To set from the Waveform screen (p. 114)
Operating Key
Procedure
1
SHEET/PAGE
Select the
[Use Ch]
page.
2
Select the number of channels to use.
(The settings of the numbers of channels for Timebase 1 and
Timebase 2 are interdependent)
Timebase 1 (Channels other than those on the 8958)
CURSOR
Move the cursor to the
[Timebase 1]
item.
F1 to F8
Select the number of channels to use.
When using logic channels:
Logic channels CH A through CH D are repre-
sented as one channel.
Timebase 2 (exclusive for channels on the 8958)
CURSOR
Move the cursor to the
[Timebase 2]
item.
F1 to F8
Select the number of channels to use.
Ch 1 to 8 and Ch 9 to 16 groups are each repre-
sented as one channel [1× 8CH]
3
Select the channels for measurement (analog/logic
inputs).
CURSOR
Move the highlight cursor to a channel to use for
measurement.
F1 to F8
Select either choice.
4
SHEET/PAGE
Select the
[Basic]
page.
5
Set the Timebase 1 (or sampling rate).
CURSOR
Move the cursor to the
[Timebase]
or
[Sampling
Speed]
item.
F1 to F8
(Switch Display: F8)
About setting ranges: "Timebase Setting: Using
the TIME/DIV Key" (p. 91)
6
Set the Timebase 2 sampling rate.
CURSOR
Move the cursor to the Timebase 2
[Sampling
Speed]
item.
F1 to F8
(Switch Display: F8)
Set the sampling rate.
The sampling rate cannot be set faster than that
of Timebase 1.
1CH, 2CH, 4CH, 8CH or 16CH
1 × 8CH, 2 × 8CH, 4 × 8CH, 8 × 8CH (default setting)
Off
No measurement.
On
Use for measurement.
Logic Channels
Channels for
measurement
1
4
: On
: Off
Input Module and
Channel Nos.
About the Number of Channels
For the Model 8861-50 case:
The total number of usable channels is
twice the number of selected channels.
However, the number of channels enabled
for use in the upper tier (Units 1 to 4+ and
Logic A to D) and lower tier (Units 5 to 8)
cannot exceed the specified number of
channels (p. 89).
When [8CH] (Timebase 1) / [8 × 8CH]
(Timebase 2) is selected with the Model
8861-50: up to eight channels (Timebase
1) / 64 channels (Timebase 2) can be used
on both the upper and lower tiers. (The to-
tal number of usable channels is 16 on
Timebase 1 plus 128 on Timebase 2.)
Summary of Contents for 8860-50
Page 2: ......
Page 12: ...Contents x...
Page 18: ...Reading this Manual 6...
Page 24: ...1 3 Interconnection and Block Diagrams 12...
Page 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Page 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Page 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Page 406: ...14 2 External I O 394...
Page 489: ......
Page 490: ......