4.2 Setting Measurement Configuration (Status Settings Screen)
91
4
Chapte
r 4
Me
asuremen
t Co
nfiguration Sett
ings
The following two setting methods are available:
• Using the operating keys
• Using the
TIME/DIV
key (settable regardless of cursor position)
The data refresh rate is not allowed to exceed the maximum sampling rate of the
input module.
Example: Using an input module with maximum sampling rate of 1 MS/s (up to
1M samples per second). 1 MS/s = 1 µs/S (1 µs sampling period)
When the
[Sampling Speed]
is set to
[50 ns/S]
, data is refreshed once each µs.
The maximum sampling rate of the input module being used can be verified on
the Config (Configuration) screen ("13.3.6 System Configuration List" (p. 374)).
Also, when sampling at different rates, if the recording time determined by the
specified recording length is shorter than the Timebase 2 sampling rate, no data
is sampled on Timebase 2.
Timebase 1
Timebase 2
Recording Length (Recording Duration)
Sampling Period
Timebase Setting: Using the TIME/DIV Key
To open the screen: Press the
SET
key
Select
with the
SUB MENU
keys
Status Settings screen
See
Screen Layout (p. 28), To set from the Waveform screen (p. 114)
Select slower
timebase
The value changes with each key press.
Select faster
timebase
Summary of Contents for 8860-50
Page 2: ......
Page 12: ...Contents x...
Page 18: ...Reading this Manual 6...
Page 24: ...1 3 Interconnection and Block Diagrams 12...
Page 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Page 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Page 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Page 406: ...14 2 External I O 394...
Page 489: ......
Page 490: ......