8.10 Fine Adjustment of Input Values (Vernier Function)
218
Fine adjustment of input voltage can be performed arbitrarily on the Waveform
screen. When recording physical values such as noise, temperature and accel-
eration using sensors, amplitude can be adjusted to facilitate calibration.
8.10 Fine Adjustment of Input Values
(Vernier Function)
Normal Display
Vernier Function Enabled
1.2 V input is displayed as 1.0 V
1.2 V
1.0 V
Press the
DISP
key to display the Waveform screen.
Using the
CURSOR
keys, move the
cursor to an item in the dialog, and
press the F keys to select it.
1
4
Select the channel to
adjust.
Adjust while viewing the waveform.
Adjustment Range: between 50 and 200% of original
waveform (magnification/compression ratio not dis-
played)
: Magnify waveform
: Compress waveform
Press the
RANGE/POSN
knob to
display the [CH SET] dialog.
2
3
• The Vernier function cannot be applied to the calculation waveforms.
• Vernier adjustments cannot be verified on printed waveforms or lists.
• When modulated by the vernier function, a marker (
or
) will be added
to the channel settings, CH information, and the channel number on the level
monitor.
Summary of Contents for 8860-50
Page 2: ......
Page 12: ...Contents x...
Page 18: ...Reading this Manual 6...
Page 24: ...1 3 Interconnection and Block Diagrams 12...
Page 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Page 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Page 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Page 406: ...14 2 External I O 394...
Page 489: ......
Page 490: ......