6.5 Pre-Trigger Settings
142
When using a [Stop] trigger at the same time
When you want to record data from a specified moment sometime after a trigger
event (that is, with a negative pre-trigger value specified), if a stop trigger event
occurs after the pre-trigger period has passed but within 2 ms after recording
starts, no waveform data is stored.
T
Pre-Trigger Setting
-x%
Recording
2 ms
Stop trigger
If a Stop trigger event occurs during this interval, no
waveform data is stored.