5.4 Converting Input Values (Scaling Function)
125
5
Chapte
r 5
Input Chan
nel Setting
s
The full span of output from a sensor can be displayed. (p. 217)
To verify correct scaling settings: Scaling Check
When appropriate numerical values have been entered,
the converted physical value is displayed.
Verify that it is converted correctly.
1
Select the
[Check]
button.
2
Select the
[Close]
but-
ton to close the dialog.
3
The [Scaling Check] dialog appears.
To reset Scaling settings
Select the
[Reset]
button.
Scaling settings are reset.
Using the Scaling and Variable functions (p. 215) in combination
At factory shipping, automatic correction of the variable function (p. 358) is set to
[On]
.
At this time, the Variable setting is altered so that it is linked to (dependent upon)
the measurement range and Scaling settings. If you want the Variable function
setting to take priority, use either of the following procedures:
• Set Scaling first, and then set the Variable function
• Set a Variable value before Scaling, and then set Scaling.
When automatic correction of the Variable function (Variable Auto Adjustment) is
disabled (Off), the Scaling and Variable settings are unlinked (independent of
one another).
Summary of Contents for 8860-50
Page 2: ......
Page 12: ...Contents x...
Page 18: ...Reading this Manual 6...
Page 24: ...1 3 Interconnection and Block Diagrams 12...
Page 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Page 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Page 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Page 406: ...14 2 External I O 394...
Page 489: ......
Page 490: ......