15.4 Recorder Function
405
15
Chap
ter 15 S
pec
ification
s
15.4 Recorder Function
Time axis
10*
1
, 20*
1
, 50*
1
, 100*
1
, 200*
1
, 500 ms/div
1, 2, 5, 10, 30, 50, 100 s/div
1, 2, 5, 10, 30 min/div
1 h/div
*1. Real-time paper recording is not available with the faster ranges (10 ms to 200 ms/
div), although waveforms can still be recorded in memory and monitored on-screen.
Up to 5000 divisions (with Model 9715-50 Memory Board installed) of waveforms
can be stored before measurement stops. Also, if the Recording Length is set to
other than [Cont], simultaneous printing is available, so waveforms can be printed
under the following conditions:
With [Cont]: Recording Length, 20 ms/div to 1 h/div
With Model 8958 16-Ch Scanner Unit installed, 50 ms/div to 1 h/div
Time axis resolution
100 points/div
Sampling period
100 ns, 1
s, 10
s, 100
s, 1 ms, 10 ms, 100 ms, 1s
(not more than 1/100th of the selected timebase)
Recording mode
Maximum and minimum recording values
Records data at a rate of 100 samples per division, with a maximum and minimum value
included in each sample.
Recording length
Fixed, User (Adjustable) or Cont (Continuous)
(setting range depends on the capacity of installed memory)
• Fixed settings
25, 50, 100, 200, 500, 1000, 2000, 5000, 10000, 20000, 50000, 100000
• Adjustable settings
1 to 160000 (in 1-div steps)
Continuous setting is not available for 10 ms to 200 ms/div timebase settings when
printing.
Timebase settings of 10 ms/div to 1 s/div are not available when printing numerical val-
ues on the Model 8995-01 A6 Printer Unit
Screen and printing
settings
1, 2, 3, 4, 6, 8 or 16 screens (printer, except on the Model 8995-01 A6 Printer Unit), can
be displayed sequentially, or selectable layout (for three divisions or more)
Sheet display (up to 32 channels displayed per sheet)
Interpolation function
Line
Recording line distinction
32 colors (four printing types)
Waveform magnification
and compression
• Time axis
× 4*, × 2*, × 1, × 1/2, × 1/5, × 1/10, × 1/20, × 1/50, × 1/100, × 1/200, × 1/500,
× 1/1000, × 1/2000, × 1/5000, × 1/10000, × 1/20000
(* Screen display only. Printing is can be x 1 or more)
• Voltage axis
× 100, × 50, × 20, × 10, × 5, × 2, × 1, × 1/2, × 1/5, × 1/10
Maximum Recording Length
[Divisions]
Installed Memory
(Words)
Other than the Model 8958 16-Ch
Scanner Unit
Model 8958 16-Ch Scanner Unit
8860-50 8861-50
Fixed
Adjustable or
Continuous
Fixed
Adjustable or
Continuous
32M
64M
5,000
5,000
1,000
1,000
128M
256M
20,000
20,000
5,000
5,000
512M
1G
50,000
80,000
20,000
20,000
1G
2G
100,000
160,000
20,000
40,000
Summary of Contents for 8860-50
Page 2: ......
Page 12: ...Contents x...
Page 18: ...Reading this Manual 6...
Page 24: ...1 3 Interconnection and Block Diagrams 12...
Page 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Page 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Page 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Page 406: ...14 2 External I O 394...
Page 489: ......
Page 490: ......