The range that can be specified by A/B cursors depends on the function.
See
"Appendix 2.4 Memory Capacity and Maximum Recording Length" (p. A35)
• With the Memory function:
The range must be within the data recorded by one measurement
• With the Recorder function:
The range can be within the data recorded by one measurement, or within the
internally recorded data that can be retraced from the end of measurement.
(see Table below)
(with [ x 1] magnification) [Divisions]
Installed Memory (Words)
8958 16-Ch Scanner Unit
8860-50
8861-50
When Uninstalled
When Installed
32M
64M
5,000
1,000
128M
256M
20,000
5,000
512M
1G
80,000
20,000
1G
2G
160,000
40,000
If the cursors do not appear on screen when the cursor type is selected
Turn the A/B knobs to display the cursors.
To compose or calculate waveforms within a specified range
After specifying the range, place the cursor on the [Exec] button and select the
type of execution by the F keys.