14.1 Connecting External Control Terminals
377
14
Chapter 14
External Control
Connecting the SYNC.OUT Jack and the EXT.SMPL Jack (BNC connector)
Connect the BNC plug of the connection cord to
the BNC jack on the instrument side.
Align the slots on the BNC plug with the guide pins on
the instrument-side jack, then push and twist the plug
clockwise until it locks.
Disconnecting BNC connectors
Twist the BNC plug counterclockwise and pull it out.
Lock
Instrument-side
guide pins
BNC plug slots
Connecting External I/O Terminals (Connector Blocks)
Cables to connect
Single strand
Multi-strand
10 mm
Connection procedure
1
2
3
• Recommended cables:
single strand diameter 0.65 mm (AWG22),
multi-strand 0.32 mm
2
(AWG22)
• Usable cables:
Single strand diameter 0.32 to 0.65 mm (AWG28 to 22),
Multi-strand 0.08 to 0.32 mm
2
(AWG28 to 22)
Strand diameter 0.12 mm or greater
• Standard insulation stripping length: 9 to10 mm
• Button operation specified tool: Flat-blade screwdriver
(tip width 2.6 mm)
1
Push in the button on the connector with a flat-
blade screwdriver or other tool.
2
With the button held in, insert the cable into the
cable connection hole.
3
Release the button.
The cable is locked.
Summary of Contents for 8860-50
Page 2: ......
Page 12: ...Contents x...
Page 18: ...Reading this Manual 6...
Page 24: ...1 3 Interconnection and Block Diagrams 12...
Page 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Page 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Page 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Page 406: ...14 2 External I O 394...
Page 489: ......
Page 490: ......