14.2 External I/O
381
14
Chapter 14
External Control
1
Connect the cables for the corresponding external input signals to the TRIG
OUT/CAL and GND terminals.
2
In the Ext Term (external terminal) Settings screen, select a setting for the
[TRIG.OUT/CAL]
terminal.
When a trigger event occurs, a pulse wave changing from the HIGH level (4.0 to
5.0 V) to the LOW level (0 to 0.5 V) is output from the TRIG OUT/CAL terminal.
Signal Output Procedure
GND is common. It can be connected to any ground.
Connection procedure:
"14.1 Connecting External Control Terminals" (p. 376)
(To open the screen: Press the
DISP
key
Press the
F7
[System]
key
Select
with the
SUB MENU
keys
Ext Term Settings Screen)
Select [Trig Out](default setting)