Short Instruction Manual
1
GENERAL DESCRIPTION
Congratulations on your new RMS MINI Logger. The MINI Logger has an internal data memory of
10,000 measured values and transmits these values continuously to the gateway and RMS soft-
ware by radio. These short instructions describe the main functions of the device.
Please read this short manual and the instruction manual on
www.rotronic.com/rms
carefully.
2
DIMENSIONS / CONNECTIONS
Top view
Rubber cap (front view)
Wall bracket
3
INSTALLATION AND SETUP
Remove the black rubber cap and pull out the PCB board. Insert the battery by following the polar-
ization sign. After the battery is inserted, the MINI-Logger starts to work.
3.1 CONNECTION OF EXTERNAL ELEMENTS (RMS-MLOG-T10-915, RMS-MADC-915-A)
Connect the external element to the terminal. The rubber cap has a hole as lead through (to be
pierced with a sharp object).
3.2 POSITIONING OF THE MINI LOGGER / MECHANICAL INSTALLATION
Choose a representative position for the installation of the Logger. Avoid disruptive infl uences
such as sunlight, heating elements, etc. The MINI Logger can be mounted easily with the wall
bracket. Screws, and anchor are provided with each MINI logger.
3.3 INTEGRATION OF THE MINI LOGGER
Log into the RMS. Select
Extras > Setup >
Devices > New > Wireless Device.
Follow the instructions given by the
software.
Rubber cap
Status LED
Pre-lead through
Button
* with wall bracket
4
LED STATUS
MINI Logger Status
LED color / interval
Explanation
Unpaired
RED flashing every 10s
Ready for pairing
Paring
ORANGE flashing
MINI Logger is detected, user input
required
GREEN flashes 3x
Pairing successful
RED flashes 3x
Pairing not successful or aborted
Connected
GREEN flashes
Status ok (LED can be switched off / on
and the interval adjusted via Software)
RED flashes 1x
Low battery <10%
RED flashes 2x
No external sensor connected or short
circuit on the sensor
Control Mode
(Press the button 1x)
GREEN flashes 1 to 4 times
Connected, 4 times = strongest signal
ORANGE flashes 1x
No wireless connection
RED flashes 1x
Low battery <10%
RED flashes 2x
No external sensor connected or circuit on
the sensor
To unpair (Press
the button >3s)
RED flashes 3x
Unpaired MINI logger must be also remo-
ved from the software
5 MAINTENANCE
The battery needs to be replaced occasionally in dependence on the logging interval. Rotronic
further recommends annual calibration of the device.
6
TECHNICAL DATA
General
Parameter
Accuracy / Information
Range of application
Temperature: -30…85 °C (RMS-MLOG-B-915 -40...85°C)
not condensing; Maximum altitude 2000 m
Storage capacity
10’000 measured values
Log-Interval
10 sec…1 h / cloud: 60 sec… 1h
Wireless
915 MHz
Range: 15...25 m indoor
IP protection
IP65 ( RMS-MLOG-B-915 = IP30 )
Software
RMS Monitoring-Software
Weight
100 g
Battery life and type
Up to 3 years (at 23 °C,
interval 1 minute) 2,7 Jahre
(RMS-MLOG-B)
RMS-BAT
MINI Logger Typ specific
RMS-MLOG-B-915
Temperature & Humidity
-40…85 °C (±0.5°C @23° C) /
0…100% RH (±3% RH @23° C)
RMS-MLOG-T-915
Temperature
-30…85 °C (±0.4°C @23° C)
RMS-MLOG-T10-915
Temperature with ext. probe Depends on the probe
RMS-MADC-915-A
Current input
0…20 mA or 4…20 mA (Shunt 110 Ohm)
±0.2mA
7
DELIVERY PACKAGE
• MINI Logger
•
Calibration certificate
• Battery
• Wall bracket
• Short instruction manual
This equipment has been tested and found to comply with the limits for a Class A digital de-
vice, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio frequency energy
and, if not installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications. Operation of this equipment in a residen-
tial area is likely to cause harmful interference in which case the user will be required to
correct the interference at his own expense
Any changes or modifications to this device not explicitly approved by manufacturer could
void your authority to operate this equipment.
83 (87*) mm
26 m
m
29 (32*) mm
87.4 mm
65 mm
3.7 mm
7.
5 m
m
14 m
m
Terminal to connect
external elements
+
–
RMS-MLOG-T10-915
RMS-MADC-915 (A)