2 Run the manipulator to the verification position. (The manipulator position
is now deviating from the verification position.)
3 Read and note current axis 4 value in degrees (example: 96.3 degrees).
4 Manually jog, only axis 4, so that the manipulator is correctly positioned to
the verification position.
5 Read and note current axis 4 value in degrees (example: 94.2 degrees).
6 Move the manipulator to its calibration position.
7 Calculate the angle difference (ie 96.3-94.2=2.1 degrees).
8 Manually jog axis 4 the calculated angle difference (-2.1). NOTE! The direction
+/- shall be the same direction as the direction used when axis 4 was manually
jogged to coincide with the verification process. In the example -2.1 degrees.
9 Make a new manual fine calibration of axis 4 with axis in -2.1 degrees position.
10 Check again against the verification position.
11 Repeat the manual tuning if needed.
12 Create a new reference if the intention is to use the reference in the future.
678
Product manual - CRB 1100
3HAC078007-001 Revision: J
© Copyright 2021-2023 ABB. All rights reserved.
6 Calibration
6.4.5 Reference calibration
Continued
Summary of Contents for CRB 1100
Page 1: ...ROBOTICS Product manual CRB 1100 ...
Page 162: ...This page is intentionally left blank ...
Page 184: ...This page is intentionally left blank ...
Page 652: ...This page is intentionally left blank ...
Page 698: ...This page is intentionally left blank ...
Page 712: ...This page is intentionally left blank ...
Page 714: ...This page is intentionally left blank ...
Page 719: ......