Spectra Optia Apheresis System Operator’s Manual
269
Supplementary Information
A
Collection Set and IDL Set: Additional Instructions for Use
Using the Needle Protector on the Inlet Needle of the Collection Set and the IDL Set
The inlet needle on the Collection Set and the IDL Set has a needle protector to protect you from accidental injury
when you remove the needle. Follow the instructions below to correctly insert and remove the inlet needle.
Inserting the inlet needle
1
Prepare the venipuncture site according to your standard operating procedure.
2
Position the needle protector away from the wings of the needle so it does not interfere with the
venipuncture.
3
Grasp the wings, remove the tip protector from the needle, and perform the venipuncture.
4
Secure the needle tubing, according to your standard operating procedure.
Removing the inlet needle
1
Release the needle tubing, according to your standard operating procedure.
2
Prepare the dressing and place it over the venipuncture site, according to your standard operating procedure.
3
Ensure that the finger hook on the needle protector points up. Slide the needle protector forward into
position under the wings of the needle.
4
Place the index finger of one hand inside the finger hook. While maintaining appropriate pressure on the
venipuncture site, pull the tubing with the other hand so that the needle slides into the needle protector.
5
Continue pulling the tubing until you hear a “click,” indicating that the needle protector is locked in place.
Once the needle is locked in the needle protector, release the finger hook while maintaining pressure on the
venipuncture site.
6
Dispose of the needle, according to your standard operating procedure.
Using the Diversion Bag to Collect a Sample From a Venipuncture
To use the diversion bag on the inlet line, perform the following steps:
1
Perform the venipuncture with the inlet needle.
2
Unclamp the inlet line.
3
Unclamp the line to the diversion bag.
4
Allow the desired volume of blood to flow into the diversion bag.
5
Clamp and then seal the line to the diversion bag. You may also remove the bag.