18
Spectra Optia Apheresis System Operator’s Manual
253
Spectra Optia System Specifications
Surges
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V in ISM bands between 0.15 MHz and 80 MHz
80% AM at 1 kHz
Power frequency
magnetic fields
IEC 61000-4-8
30 A/m
Voltage dips
IEC 61000-4-11
0% U45; 0.5 cycle
At phase angles 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
UT 100 V and 230 V
0% UT; 1 cycle
70% UT; 25 cycles
Single phase: at 0°
UT 100 V and 230 V
Voltage interruptions
IEC 61000-4-11
0% UT; 250 cycles
UT 100 V and 230 V
Note: UT is the voltage of the main alternating-current power supply before use of the test level.
Electromagnetic environment–guidance for IEC 61000-4-11
If the user of the Spectra Optia system is aware of frequent power interruptions at his or her facility and requires continued operation during
mains power interruptions, it is recommended that the Spectra Optia system be powered from a suitably rated uninterruptible power supply.
Table 18-29:
Test specification for ENCLOSURE PORT IMMUNITY to RF wireless comm. equip. 60601-1-2 (4th Ed.)
Test
frequency
(MHz)
Band (MHz)
Service
Modulation
Maximum
power (W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385
380 to 390
TETRA 400
Pulse modulation 18 Hz
1.8
0.3
27
450
430 to 470
GMRS 460
FRS 460
FM ± 5 kHz
deviation 1
kHz sine
2
0.3
28
710
704 to 787
LTE Band 13, 17
Pulse modulation
217 Hz
0.2
0.3
9
745
780
810
800 to 960
GSM 800/900
TETRA 800
iDEN 820
CDMA 850
LTE Band 5
Pulse modulation
b) 18 Hz
2
0.3
28
870
930
Table 18-28:
Immunity tests and test levels 60601-1-2 (4th Ed.) (continued)
Immunity Test
Test Level IEC 60601