Configuring the Spectra Optia System
50
Spectra Optia Apheresis System Operator’s Manual
Procedure
Table 4-2 shows the operating parameters to configure for procedures.
Table 4-2:
Procedure configuration parameters and setting options
Tab
Parameter
Description
Setting Options
Procedure
Height units
Unit of measure for the patient’s height.
• ft, in (feet, inches)
• cm (centimeters)
Weight units
Unit of measure for the patient’s weight.
• lb (pounds)
• kg (kilograms)
Pressure alarm limit:
Inlet (mmHg)
Limit at which an inlet pressure alarm occurs.
−
250 to
−
100
(Default:
−
250)
Pressure alarm limit:
Return (mmHg)
Limit at which a return pressure alarm occurs.
200 to 400
Default: 400
Custom prime
recommendation (% TBV)
Percentage of the patient’s total blood volume (TBV) or
RBC volume in the tubing set at which the system
displays a screen recommending a custom prime.
• 10
• 15
Default: 10
AC container: Notification
Indication of whether the system will display a
notification when the AC container is nearly empty.
• Yes
• No
Default: Yes
AC container: Volume (mL)
Volume of AC in the AC container used during the
procedure.
250 to 1000
Default: 1000