18
Spectra Optia Apheresis System Operator’s Manual
243
Spectra Optia System Specifications
Physical
Restrictions
Device is not for use in an explosive
atmosphere.
Shipping temperature
−
29 °C to 60 °C (
−
20 °F to 140 °F)
Shipping humidity
8% to 90%
Relative humidity (RH), non-condensing
Storage temperature
0 °C to 60 °C (32 °F to 140 °F)
Storage humidity
8% to 80%
Relative humidity (RH), non-condensing
Table 18-3:
Environmental: Seal Safe system
Characteristics
Performance
Conditions
Ambient operating temperature
(sealer head with RF cable)
15.5 °C to 27.7 °C (60 °F to 82 °F)
Ambient operating humidity
8% to 80%
Relative humidity (RH), non-condensing
Cleaning
Components can be cleaned and disinfected.
Device is not damaged by cleaning and
disinfecting with the compatible cleaning
solutions and disinfecting solutions, according
to the instructions described in Chapter 17 of
this manual.
Fluid spillage
The unit is safe to use if fluid is spilled on the
sealer head or on the RF cable.
Table 18-4:
Environmental: Tubing sets
Characteristics
Performance
Conditions
Storage temperature
• Long term storage range: 0 °C to 35 °C
(32 °F to 95 °F)
• Permitted excursions:
•
−
29 °C to 0 °C (
−
20 °F to 32 °F) for up to
72 hours
• 35 °C to 50 °C (95 °F to 122 °F) for up to
6 weeks
Storage humidity
• 0% to 75%
• Permitted excursions: Up to 85% RH ± 5%
for up to 72 hours
Relative humidity (RH), non-condensing
Table 18-5:
Physical: Spectra Optia system
Characteristics
Performance
Conditions
Floor space required
0.43 m
2
(4.6 ft
2
)
Applies to a floor slope of less than 5°
Minimum clearance
30 cm (11.8 in) around the perimeter of the
device
Table 18-2:
Environmental: Spectra Optia device (continued)
Characteristics
Performance
Conditions
SpectraOptiaOpsManual.book Page 243 Thursday, May 17, 2018 11:24 AM