Spectra Optia Apheresis System Operator’s Manual
59
Configuring the Spectra Optia System
4
Configuring the Use of a Blood Warmer
This section contains information about configuring operating parameters for the use of a blood warmer.
Parameters for configuring the use of a blood warmer can vary according to the procedure you are performing.
Table 4-16 shows all available parameters and setting options.
Table 4-16:
Parameters and setting options to configure for using a blood warmer
Tab
Parameter
Description
Setting Options
Blood
Warmer
Return line
Indication that a blood warmer will be connected to the
return line.
You must prime the blood warmer tubing set before you
connect the patient.
• Yes
• No
Default: No
Replace line
Indication that a blood warmer will be connected to the
replace line during a procedure using an Exchange Set.
• Yes
• No
Default: No
Tubing set (mL)
Volume of the blood warmer tubing set on the return
line.
The system uses this information to adjust the
calculation for the extracorporeal volume, the minimum
volume required for a custom prime, and the volume of
saline required for rinseback.
1 to 100
Default: 40