Bone Marrow Processing (BMP) Procedures
198
Spectra Optia Apheresis System Operator’s Manual
Entering and Changing Data During the Run
The Spectra Optia system requires that you enter specific data, such as BM data, to perform a BMP procedure. The
system also allows you to change numerical values for certain operating parameters during the run. Table 14-2
shows data entry ranges for the operating parameters according to the screen on which each data entry button
appears. To change or enter data, touch the button for the parameter that you want to change. A data entry pad
appears to enable you to enter a number. The data entry pad displays either the established range for the parameter
or the calculated range, if the range changes based on related data. Also, the range for Caution status is displayed in
yellow, if the data you enter could cause the system to operate in Caution status.
Table 14-2:
Data entry ranges for BMP procedures
Screen
Parameter
Range
BM data
BMV (mL)
300 to 3000
Hct (%)
10 to 50
Options
Rinseback
• Yes
• No
Default: Yes
Saline rinse
• Yes
• No
Default: No
Collect plasma
• Yes
• No
Run values
BM processed (mL)
BMV multiplied by 4 to 10
Default: BMV multiplied by the
number of BM cycles.
(Minimum volume depends on the
volume of RBC in the BMP bag.)
BM cycles
4 to 10
(Minimum number of cycles depends
on the volume of RBC in the BMP
bag.)
Inlet flow rate (mL/min)
20 to 142
Plasma volume (mL): Target
0 to 500
Collect flow rate (mL/min)
1 to 5
Default: inlet pump flow rate
multiplied by 0.0167
Collect volume (mL): Target
Up to 500
Main run
Packing factor
1 to 20
Default: 20
Inlet flow rate (mL/min)
20 to 142
Collection status
Collection preference
10 to 90
Default: 50 after the interface is
established