Configuring the Spectra Optia System
58
Spectra Optia Apheresis System Operator’s Manual
Platelet Depletion (PLTD)
Table 4-14 shows the operating parameters to configure for PLTD procedure.
Bone Marrow Processing (BMP)
Table 4-15 shows the operating parameters to configure for a BMP procedure.
Table 4-14:
Parameters and setting options to configure for a PLTD procedure
Tab
Parameter
Description
Setting Options
PLTD
Run target: Whole blood
processed (mL)
Amount of the patient’s blood to be processed. Does not
include AC.
If you enter a value for this parameter, the system
automatically sets the targets for the run time and the
TBV processed to zero.
100 to 50000
Default: 0
Run target: Run time (min)
Time required to complete the run. The run starts when
the patient is connected and ends when the run targets
have been attained.
If you enter a value for this parameter, the system
automatically sets the targets for the whole blood
processed and the TBV processed to zero.
40 to 480
Default: 0
Run target: TBV processed
Number of the patient’s total blood volumes to be
processed.
If you enter a value for this parameter, the system
automatically sets the targets for the whole blood
processed and the run time to zero.
0.5 to 5
Default: 1.5
AC infusion
rate (mL/min/L TBV)
Rate at which AC is infused to the patient.
The configured rate represents the maximum rate at the
start of the run.
0.8 to 1.2
Default: 0.8
Inlet:AC ratio (__:1)
Ratio of the inlet pump flow rate to the AC pump flow
rate.
The ratio configured represents the ratio at the start of
the run.
6 to 15
Default: 10
Custom replacement
fluid (%)
Percentage of citrate in the custom replacement fluid.
0 to 25
Default: 0
Table 4-15:
Parameters and setting options to configure for BMP procedure
Tab
Parameter
Description
Setting Options
BMP
Inlet flow rate (mL/min)
Maximum flow rate at which the inlet pump pumps
bone marrow from the BMP bag into the tubing set
during the run.
20 to 142
Default:120
Collect plasma
Indication that plasma will be collected.
• Yes
• No
Default: No