Supplementary Information
274
Spectra Optia Apheresis System Operator’s Manual
Printing Procedure Data Reports
The Spectra Optia system stores procedure data reports for up to one hundred procedures. Perform the steps below
to connect a printer or a computer to the Spectra Optia system and to print a copy of a report.
Connecting a Printer to the Spectra Optia System
Ensure that the printer meets the following requirements before connecting it to the system:
•
Printer must be network compatible. The system cannot communicate with Universal Serial Bus (USB) or
parallel printers.
•
Printer must be compatible with Adobe
®
PostScript
®
3
™
page description language. The system does not
communicate properly with printers that are only compatible with earlier versions of the language.
•
Printer must have a configurable Internet Protocol (IP) address. The address you assign to the printer must
begin with 172.21 to correspond with the system’s address.
•
Printer must be able to use transmission control protocol (TCP) port number 9100 for Local Area Network
(LAN) printing.
•
Ethernet cable used to connect the printer to the system must be a crossover cable.
Perform the following steps to connect your Spectra Optia system to a printer:
1
Connect one end of the crossover cable to the printer and the other end of the cable to the Ethernet port on
the Spectra Optia system. The Ethernet port is located on the bottom right side of the back of the system next
to the connection for the power cord.
2
Configure the parameters and options on the report configuration screen to use a printer to print a copy of
the report. See Table 4-3 on page 51 for descriptions of the parameters and options.
3
Touch
Test
on the connection button to verify the connection between the system and the printer.
For additional information about connecting a printer to the Spectra Optia system, contact your Terumo BCT
service representative.
Note:
If you are connecting the system to the printer via a network device, such as a router or a switch, use a
standard Ethernet cable.
SpectraOptiaOpsManual.book Page 274 Thursday, May 17, 2018 11:24 AM