Spectra Optia Apheresis System Operator’s Manual
79
Selecting Procedure Options
6
Selecting and Performing a Saline Rinse
The system prompts you to perform the saline rinse after it primes the tubing set and before you connect the
patient. If you perform a saline rinse, you need the following supplies:
•
New container of saline that contains at least 250 mL.
•
Empty bag of sufficient volume to hold the saline that is used for the saline rinse.
•
Appropriate connector to connect the return line to the empty bag.
The inlet pump flow rate during a saline rinse depends on the tubing set you use. Ensure that the access to the bag
on the return line can accommodate the flow rate. The inlet pump flow rates for the tubing sets are as follows:
•
Exchange Set: 120 mL/min
•
Collection Set: 90 mL/min
•
IDL Set: 90 mL/min
Selecting the saline rinse option
To select the saline rinse option, touch the saline rinse button until
Yes
appears on the button. Resume the
procedure.
Starting the saline rinse
1
When the screen appears with the message “Prepare for saline rinse,” perform the steps that appear on the
screen:
a. Spike the new container of saline, and squeeze the drip chamber to ensure that saline drips into the
chamber.
b. Connect the empty bag to the return line.
2
Touch
Continue
.
3
Perform the steps that appear on the screen:
a. Wait 15 minutes to allow any residual ethylene oxide in the tubing set to diffuse into the saline.
b. Open the inlet saline line and the return saline line.
c. Unclamp the return line and flush the line to the access site with at least 50 mL of saline.
d. Close the return saline line.
4
Touch
Start Saline Rinse
. The status of the saline rinse appears on the screen.
5
To stop the saline rinse before it is complete, touch
Cancel Saline Rinse
. A screen appears on which you
must confirm your decision to cancel the saline rinse. Touch
End Saline Rinse
, and then touch
Confirm
.
Note:
The system uses 250 mL of saline for a saline rinse. If you intend to perform a saline rinse more than
once, you need additional saline.