5
Spectra Optia Apheresis System Operator’s Manual
69
Preparing to Perform a Procedure
9
Starting with the connector (Figure 5-6, number 1), insert the channel into the groove in the filler, finishing
with the inlet port (Figure 5-6, number 2). Run your finger over the groove and push down any section of
the channel that is not completely inserted in the groove. The channel must sit flush with the groove.
Figure 5-6:
Channel pressed into the groove (Exchange Set shown here)
10
If you are using a Collection Set, put the chamber (Figure 5-7, number 3) into the chamber bracket. Ensure
that the chamber sits behind the clip at the opening of the bracket, so that the clip retains the chamber in the
bracket. Put the collect line to the chamber behind the lip of the optical reference (Figure 5-7, number 4), so
that the line does not block the optical reference.
Figure 5-7:
Collect line correctly positioned behind the lip of the optical reference
1
2
4
3