Spectra Optia Apheresis System Operator’s Manual
113
Therapeutic Plasma Exchange with a Secondary Plasma Device (TPE-SPD) Procedures
8
a. Touch the pause button to pause the pumps. An alarm screen appears.
b. Clamp the line to the waste bag and unclamp the line to the treated plasma bag to allow the system to
begin filling the treated plasma bag.
c. Touch
Continue
to restart the pumps.
d. Once the system processes an addition 50 mL of plasma, clamp the replace line to the albumin container
and unclamp the second replace line to the treated plasma bag. The system begins returning the treated
plasma to the patient.
SpectraOptiaOpsManual.book Page 113 Thursday, May 17, 2018 11:24 AM