Therapeutic Plasma Exchange with a Secondary Plasma Device (TPE-SPD) Procedures
106
Spectra Optia Apheresis System Operator’s Manual
6
Touch
Start Prime
. The system starts priming the plasma device.
• To pause the prime, touch
Pause Prime
.
• To restart the prime, touch
Start Prime
.
The screen displays the progress of the prime.
• Current Prime indicates the volume of saline that has been processed for the current prime.
• Total Prime indicates the total volume of saline that was processed for priming the plasma device during
the current procedure.
7
After the volume for the prime has been processed, perform one of the following steps:
• To process additional volume, repeat steps 5 and 6.
• If you are finished priming the plasma device, touch
Prime Complete
.
8
Follow the instructions on the screen to clamp the replace line, spike the second replace line into the treated
plasma bag, unclamp the second replace line, and connect the treated plasma bag to the outlet end of the
plasma device.
9
Touch
Confirm
.
Saline prime: No
1
If you intend to perform a prime divert, follow the instructions on the screen to connect a waste bag to the
outlet end of the plasma device.
2
Touch
Confirm
.
3
Follow the instructions on the screen to spike the second replace line into the treated plasma bag, and to
connect the treated plasma bag to the outlet end of the plasma device.
4
Touch
Confirm
.
Priming the Inlet Line and the Return Line
1
Follow the instructions on the screen to prime the inlet line, the return line, and the blood warmer tubing set,
if used.
2
Touch
Confirm
. The screen appears with instructions for connecting the patient and starting the run.
Warning:
When connecting a blood warmer tubing set to the return line, ensure that the tubing connection is
tight. Put the luer connection no higher than 20 in (50 cm) above the return access to prevent the possibility of air
entering the tubing.
Warning:
If you are using a blood warmer on the return line, ensure that you completely prime the blood warmer
tubing set to remove all the air in the set before you connect the patient.