Therapeutic Plasma Exchange with a Secondary Plasma Device (TPE-SPD) Procedures
102
Spectra Optia Apheresis System Operator’s Manual
Bolus
Volume (mL)
10 to a volume that 25% of
the patient’s TBV
Flow rate (mL/min)
10 to 120
Run targets
Plasma volumes treated
0.5 to 4
Run time (min)
40 to 480
Plasma treated (mL)
Depends on the patient’s plasma
volume.
Treated plasma
Volume (mL)
5 to 2000
Return flow rate (mL/min)
2 to 100
Plasma device rinse
Volume (mL)
5 to 2000
Flow rate (mL/min)
1 to 100
Rinseback
Return flow rate (mL/min)
2 to 100
Default: inlet flow rate that was in
use when the run ended or
100 mL/min, whichever is slower
Table 8-2:
Data entry ranges for TPE-SPD procedure parameters (continued)
Screen
Parameter
Range
SpectraOptiaOpsManual.book Page 102 Thursday, May 17, 2018 11:24 AM