Spectra Optia Apheresis System Operator’s Manual
245
Spectra Optia System Specifications
System Components
Centrifuge
Pumps
Table 18-9:
Centrifuge speed and g-force
Characteristics
Performance
Conditions
Centrifuge speed
• Standard filler: up to 3,000 rpm
• IDL filler: 550 to 2,500 rpm
Maximum g-force in the channel 1,200 g
Device is operating at maximum speed.
Table 18-10:
Pump flow rates
Characteristics
Performance
Conditions
AC pump
Up to 12 mL/min
When a patient is connected
Inlet pump
• MNC collection procedures: Up to 125 mL/min
• All other procedures: Up to 142 mL/min
Plasma pump
• MNC collection procedures: Up to 125 mL/min
• All other procedures: Up to 142 mL/min
Operator can enter a pump flow rate
from within this range during the
procedure.
Collect pump
• MNC collection procedures: 0.5 mL/min to
10 mL/min
• During accumulation phase: Up to 3 mL/min
• During collection phase: Up to 7 mL/min
Operator can enter a pump flow rate
from within this range during the
procedure.
• CMNC procedures: 0.5 mL/min to 10 mL/min
• PMN collection, WBCD, PLTD procedures:
0.5 mL/min to 25 mL/min
• BMP procedures: 1 mL/min to 5 mL/min
Replace pump
• TPE, TPE-SPD, RBCX procedures: Up to 150 mL/min
• During administration of a bolus: 10 to 120 mL/min
• Normal operating conditions
• Operator can enter a pump flow rate
from within this range for the
administration of a bolus.
Return pump
• All procedures: Up to 295 mL/min
• During rinseback: 2 mL/min to 100 mL/min
• Normal operating conditions
• Operator can enter a pump flow rate
from within this range for rinseback.