Spectra Optia Apheresis System Operator’s Manual
151
Continuous Mononuclear Cell Collection (CMNC) Procedures
11
Volume (mL)
100 to 400
(Volume should be at least 300 mL
for the patient to receive the full
benefit.)
Run values
AC infusion rate (mL/min/L TBV)
0.2 to 2.5
(Caution status applies if the entry
exceeds 1.2.)
Inlet:AC ratio (__:1)
2 to 50
(Warning appears if the entered ratio
is not within the recommended range
of 6:1 to 15:1.)
Whole blood processed (mL)
Patient’s TBV multiplied by 1 to 10
Run time (min)
40 to 480
TBV processed
1 to 10
Inlet flow rate (mL/min)
5 to 142
Plasma volume: Target (mL) in plasma bag
0 to the patient’s plasma volume
minus the current collect volume and
the current volume of plasma in the
collection bag
(Sum of the total plasma volume and
the collect volume cannot exceed
100% of the patient’s total plasma
volume.)
Collect flow rate (mL/min)
0.5 to 10
Default:
• 1.0 if the inlet flow rate is equal to
or greater than 20
• Inlet flow rate multiplied by 0.05 if
the inlet flow rate is less than 20
Collect volume: Target (mL)
5 to the patient’s plasma volume
minus the current volume of plasma
in the plasma bag and in the
collection bag
Plasma in collection bag: Target (mL)
0 to the patient’s plasma volume
minus the current collect volume and
the current volume of plasma in the
plasma bag
(Sum of the total plasma volume and
the collect volume cannot exceed
100% of the patient’s total plasma
volume.)
Main run
Packing factor
1 to 20
Default: 4.5
Inlet flow rate (mL/min)
5 to 142
Table 11-2:
Data entry ranges for CMNC procedure parameters (continued)
Screen
Parameter
Range