17
Spectra Optia Apheresis System Operator’s Manual
237
Maintaining the Spectra Optia System
Reassembling the sealer head
1
Ensure that the spring is in place in the head.
2
Locate the grooves on both sides of the head. Align the jaw with the grooves and slide it down the sealer head
until it touches the spring (Figure 17-6). Ensure that the spring fits into the indentation in the jaw.
Figure 17-6:
Aligning the jaw with the grooves
3
Use your thumb to slightly depress the jaw. Ensure that the spring stays in place.
4
Replace the lever by tipping it at a 45° angle and positioning the lever pivots behind and below the pivot
slots (Figure 17-7).
Figure 17-7:
Replacing the lever
5
Push down the lever to engage the lever pivots while simultaneously depressing the jaw with your thumb.
6
Release your thumb. The lever should fully engage and secure the jaw on the sealer head.
7
Squeeze the lever to ensure that the head mechanically functions. If the jaw does not completely close, check
the position of the spring.
8
Attach a new splash guard, if required, according to the instructions “Attaching a new splash guard” below.
9
Reconnect the RF cable to the sealer head and put the sealer head back into the holder.
SpectraOptiaOpsManual.book Page 237 Thursday, May 17, 2018 11:24 AM