Spectra Optia Apheresis System Operator’s Manual
205
Bone Marrow Processing (BMP) Procedures
14
Changing the Target Plasma Volume
The system collects the target plasma volume after the other run targets have been attained. The plasma valve
moves into the collect position, and the system pumps plasma into the plasma bag. A message appears on the
message bar indicating that the plasma collection is in process. When the plasma collection is complete, the plasma
valve moves into the return position.
If you decide to collect plasma and you did not previously configure the procedure for plasma collection or select
the option to collect plasma, perform the following steps:
1
Touch the target button for the plasma bag, and enter the desired volume. The screen appears instructing you
to prepare the tubing set for plasma collection.
2
If you are using an IDL Set with catalog number 10300, spike the plasma bag to the tubing set, or connect
the plasma bag according to the instructions under “Using a sterile tubing welder to connect the plasma bag
to the replace line” on page 196. The run values screen appears.
3
Touch
Confirm
. The run targets screen appears and the system collects that target plasma volume.
Concentrating the RBC in the BMP Bag
To concentrate the RBC in the BMP bag, remove as much plasma as possible from the BMP bag. Follow the
instructions under “Changing the Target Plasma Volume” above and enter the maximum volume of plasma
available to collect. The maximum volume is displayed in the volume range on the data entry pad. This volume is
based on the volume of plasma required to maintain a maximum Hct of 65% in the BMP bag.