![Terumo BCT Spectra Optia Apheresis System Скачать руководство пользователя страница 179](http://html1.mh-extra.com/html/terumo-bct/spectra-optia-apheresis-system/spectra-optia-apheresis-system_operators-manual_1090434179.webp)
Spectra Optia Apheresis System Operator’s Manual
165
Granulocyte (PMN) Collection Procedures
12
Entering and Changing Data During the Run
The Spectra Optia system requires that you enter specific data, such as patient data, to perform a PMN collection
procedure. The system also allows you to change numerical values for certain operating parameters during the run.
Table 12-2 shows data entry ranges for the operating parameters according to the screen on which each data entry
button appears. To change or enter data, touch the button for the parameter that you want to change. A data entry
pad appears to enable you to enter a number. The data entry pad displays either the established range for the
parameter or the calculated range, if the range changes based on related data. Also, the range for Caution status is
displayed in yellow, if the data you enter could cause the system to operate in Caution status.
Table 12-2:
Data entry ranges for PMN collection procedures
Screen
Parameter
Range
Patient data
Sex
• Male
• Female
Height:
• ft
• in
• cm
• 1 to 8
• 0 to 11
• 30 to 244
Weight:
• lb
• kg
• 5 to 500
• 2 to 227
TBV (mL)
300 to 15000
Hct (%)
10 to 70
Options
Rinseback
• Yes
• No
Default:
• Yes if a custom prime is not
selected
• No if a custom prime is selected
Custom prime
• Yes
• No
Default: No
HES
Selection determines the default packing factor.
• Yes: Default packing factor is 1.6
• No: Default packing factor is 4.5
• Yes
• No
Saline rinse
• Yes
• No
Default: No
Blood warmer: Return line
• Yes
• No
Blood warmer: Tubing set volume (mL)
1 to 100
Custom prime
Fluid type
• RBC
• Plasma
• Albumin
SpectraOptiaOpsManual.book Page 165 Thursday, May 17, 2018 11:24 AM