Spectra Optia Apheresis System Operator’s Manual
51
Configuring the Spectra Optia System
4
Report
Table 4-3 shows the operating parameters to configure for the report.
Table 4-3:
Report configuration parameters and setting options
Tab
Parameter
Description
Setting Options
Report
Device
Device that is used to deliver a copy of the report.
If you select “Printer,” you must also select a paper size.
• No Device
• Printer
• Computer
Default: No Device
Paper
Size of paper that is used to print the report if “Printer”
is the device selected.
• Letter
• A4
Default: Letter
IP address
IP address for the device that is used to print the report.
The format for the address is XXX.XX.XXX.XXX. The first
five digits must be 172.21 to match the system’s
address.
Device-specific
Port
Port number for the device that is used to print the
report. The format for the number is XXXX. It is
recommended that you use the default number.
100 to 9999
Default: 9100
Connection
Test that is used to verify that a device for printing a
report is connected to the system.
Touch
Test
to verify the connection. If the system can
verify the connection, “Passed” appears on the button.
If the system cannot verify the connection, “Failed”
appears.
Not configurable
Report
Type of procedure report that you want to view or print.
There are two types of reports:
• Summary: Single-page report including the system
serial number, the date of the procedure, the patient
data, the fluid balance, and the initial and final run
values)
• Detail: Summary report plus additional procedure
details, which are reported according to time
configured for the report update interval.
• Summary
• Detail
Default: Summary
Report update
interval (min)
Interval at which procedure information appears on the
report.
10 to 60
Default: 15
SpectraOptiaOpsManual.book Page 51 Thursday, May 17, 2018 11:24 AM