Spectra Optia Apheresis System Operator’s Manual
247
Spectra Optia System Specifications
Monitor and touch screen
Default alarm point
• Inlet pressure sensor:
–250 mmHg ± 30 mmHg
• Return pressure sensor:
+400 mmHg ± 37 mmHg
Pressure accuracy
± 6% of the sensor reading or ± 20 mmHg,
whichever is greater
Table 18-16:
RBC detector
Characteristics
Performance
Conditions
RBC detection level
> 1.5% hematocrit
• TPE procedures: System automatically
diverts fluid away from the remove bag
when it detects RBC.
• MNC collection procedures: System initiates
a collection phase when it detects RBC
exiting the chamber.
Table 18-17:
Low-level and high-level reservoir sensors
Characteristics
Performance
Conditions
Pump stroke volume between
sensors
• 49 mL ± 9 mL
• Sensors detect foam as air.
Level sensor requires a load force of > 5 lb
(2.3 kg) each (> 10 lb [4.5 kg] total).
Alarm occurrence
When fluid is not detected at the low-level
sensor, the system stops the pumps before air
enters the line.
System performs a duplicate (redundant)
safety check. If condition is not resolved after
a specific number of tries, operator must
disconnect patient.
Table 18-18:
Return line air detector
Characteristics
Performance
Conditions
Alarm occurrence
Alarm is generated if the system calculates
that the patient might have more than 1 mL of
air in circulation.
Normal operating conditions
Table 18-19:
Monitor and touch screen
Characteristics
Performance
Conditions
Monitor with graphical user
interface (GUI) touch screen
• Monitor holds a touch screen that provides a
tactile operator interface.
• Monitor has a speaker for sounding
procedure and alarm tones, alarm lights,
and two external dedicated buttons with
icons (stop button and pause button).
Screen is touched to cause a change in
resistance.
Size
Measures 26.7 cm (10.5 in) diagonally
Table 18-15:
Inlet and return pressure sensors (continued)
Characteristics
Performance
Conditions