Installing and Transporting the Spectra Optia System
44
Spectra Optia Apheresis System Operator’s Manual
Transporting the Spectra Optia System
The Spectra Optia system can be transported manually or in a vehicle. It is not necessary to recalibrate the system
after transport; the system is ready for use.
Preparing the Spectra Optia System for Transport
1
Unplug the Spectra Optia system and secure the power cord in the power cord holder.
2
Press the IV pole release button located on the left side of the front panel and lower the IV pole.
3
Lower the monitor and lay it over the front panel (Figure 3-1).
Figure 3-1:
Monitor lowered over the front panel
4
Open the centrifuge door and confirm that the filler latch on the filler is lowered and locked. Close the
centrifuge door.
5
Remove the following components and transport them separately to avoid damage to the system and the
components:
• Blood warmer and accessory bracket, if used.
• Seal Safe system sealer head and cable.
6
If you are transporting the system in a vehicle, cover the system with a heavy-duty cover or a shipping
blanket to protect the system.
Caution:
To avoid possible damage to the Seal Safe system, always disconnect the sealer head and the
RF cable before transporting the Spectra Optia system.