Maintaining the Spectra Optia System
234
Spectra Optia Apheresis System Operator’s Manual
Covers on the AIM system lights
1
Wipe the covers, using water or a disinfecting solution.
2
After disinfecting, wipe the covers again using water to remove any residue that was left by the disinfecting
solution.
3
Dry the covers, using a gauze pad or a soft, lint-free cloth.
Filler
1
Follow the instructions on page 64 to remove the filler from the centrifuge.
2
Wipe the filler, using a mild detergent or a disinfecting solution. Use a cotton swab to clean the groove of
the filler and a dry cotton swab or alcohol pad to remove excess fluid.
3
Wipe the surfaces of the aperture plate, using water to remove any residue that was left by the cleaning
solution or disinfecting solution.
4
Dry the aperture plate, using a gauze pad or a soft, lint-free cloth.
5
Allow the remaining surfaces of the filler to air-dry completely.
6
Follow the instructions on page 65 to re-install the filler on the centrifuge.
Seal Safe System: Jaw Cavity
Examine the sealer head after each procedure. If you see dirt or fluid in the jaw cavity, follow the instructions below
to remove the residue. It is important that the jaw cavity be kept clean. Residue left in the cavity can impair the
performance of the device.
1
Disconnect the sealer head from the RF cable.
2
Remove and discard the splash guard, if used.
Caution:
Use only a gauze pad, a lint-free cloth, or a wipe when cleaning or disinfecting the touch screen, the
covers on the AIM system lights in the centrifuge, and the aperture plate on the filler. Use of an abrasive brush,
scrub material, or a sharp object can damage the surface of the components.
Note:
If necessary, you can rinse the filler with water or mild detergent before disinfection to loosen any
residue left by a spill.
Warning:
Disconnect the sealer head from the RF cable before cleaning to avoid receiving a serious radio
frequency (RF) burn during the cleaning process.