White Blood Cell Depletion (WBCD) and Platelet Depletion (PLTD) Procedures
182
Spectra Optia Apheresis System Operator’s Manual
Custom prime
Fluid type
• RBC
• Plasma
• Albumin
RBC unit Hct (%)
25 to 80
Inlet flow rate (mL/min)
5 to 70
Volume (mL)
100 to 400
(Volume should be at least 300 mL
for the patient to receive the full
benefit.)
Run values
AC infusion rate (mL/min/L TBV)
0.2 to 2.5
(Caution status applies if the entry
exceeds 1.2.)
Inlet:AC ratio (__:1)
2 to 50
(Warning appears if the entered ratio
is not within the recommended range
of 6:1 to 15:1.)
Whole blood processed (mL)
Patient’s TBV multiplied by 0.5 to 5
Run time (min)
40 to 480
TBV processed
0.5 to 5
Target fluid balance (mL or %)
70% to 125% of the patient’s TBV
(Caution status applies if the entry
exceeds ± 15% of the patient’s TBV.)
Replacement fluid:
• Fluid type
• No fluid
• Plasma
• Saline/Albumin
• RBC
• Custom
• Average Hct of RBC (%)
• Citrate content (%) of custom fluid
• 10 to 80
• 0 to 25
• Volume (mL)
10 to patient’s TBV
Inlet flow rate (mL/min)
5 to 142
Collect flow rate (mL/min)
0.5 to 25
Default:
• WBCD: depends on the patient’s
WBC count
• PLTD: depends on the patient’s
platelet count
Collect volume: Target (mL)
5 to the patient’s plasma volume
Table 13-2:
Data entry ranges for WBCD and PLTD procedures (continued)
Screen
Parameter
Range
SpectraOptiaOpsManual.book Page 182 Thursday, May 17, 2018 11:24 AM