Continuous Mononuclear Cell Collection (CMNC) Procedures
152
Spectra Optia Apheresis System Operator’s Manual
Collect flow rate (mL/min)
0.5 to 10
Collect volume (mL)
5 to the patient’s plasma volume
minus the current volume of plasma
in the plasma bag and in the
collection bag.
AC infusion rate (mL/min/L TBV)
0.2 to 2.5.
(Caution status applies if the entry
exceeds 1.2.)
Inlet:AC ratio (__:1)
2 to 50
(Warning appears if the entered ratio
is not within the recommended range
of 6:1 to 15:1.)
Collection status
Collection preference
10 to 90
Default: 50 after the interface is
established
Run targets
Plasma bag (mL)
0 to the patient’s plasma volume
minus the current collect volume and
the current volume of plasma in the
collection bag
(Sum of the total plasma volume and
the collect volume cannot exceed
100% of the patient’s total plasma
volume.)
Run time (min)
40 to 480
Whole blood processed (mL)
Patient’s TBV multiplied by 1 to 10
TBV processed
1 to 10
Collection bag: Plasma (mL)
0 to the patient’s plasma volume
minus the current collect volume and
the current volume of plasma in the
plasma bag
(Sum of the total plasma volume and
the collect volume cannot exceed
100% of the patient’s total plasma
volume.)
Collection bag: Collect (mL)
5 to patient’s plasma volume minus
the current volume of plasma in the
plasma bag and in the collection bag.
Rinseback
Return flow rate (mL/min)
2 to 100
Default: inlet flow rate that was in
use when the run ended or
100 mL/min, whichever is slower
Table 11-2:
Data entry ranges for CMNC procedure parameters (continued)
Screen
Parameter
Range