Ending the Run and Completing the Procedure
214
Spectra Optia Apheresis System Operator’s Manual
Removing the Tubing Set
Once you finish the procedure and raise the cassette, perform the following steps to remove the tubing set:
1
Open the centrifuge door.
2
Remove the upper collar from the collar holder by grasping the lines above and below the collar and pulling
the lines downward.
3
Remove the upper and lower bearings from the bearing holders.
4
If you are using a Collection Set, remove the chamber from the bracket.
5
Gently pull the channel from the filler.
6
Push in the locking pin on the centrifuge collar holder, and remove the collar from the holder by grasping
the tubes above the collar and pulling upward.
7
Push the filler latch pin toward the center of the centrifuge, and raise the filler latch.
8
Fold the channel in half, and pull the channel through the loading port and out of the centrifuge chamber.
9
Lower the filler latch.
10
Close the centrifuge door.
11
Remove the lines from the fluid detectors.
12
Remove any bags from the IV pole.
13
Press the latch on the upper right corner of the cassette tray, and lift the cassette from the tray.
14
Discard the tubing set, according to your standard operating procedures.
Note:
If you do not perform rinseback, the channel of the tubing set will be full of fluid when you unload the set. After you
raise the cassette, empty the channel by putting the cassette and the bags at a level below the channel, allowing the fluid in
the channel to drain into the vent bag.
SpectraOptiaOpsManual.book Page 214 Thursday, May 17, 2018 11:24 AM