Maintaining the Spectra Optia System
230
Spectra Optia Apheresis System Operator’s Manual
Cleaning and Disinfecting the Spectra Optia System
This section contains instructions for cleaning and disinfecting the Spectra Optia system. Follow the instructions
for cleaning the system to remove dirt and other substances that could impair the function of the system. Follow
the instructions for disinfecting the system to eliminate potentially infectious substances. Always confirm that the
system is clean and undamaged before use.
Compatible Cleaning Solutions
The cleaning solutions listed below are compatible for use on the Spectra Optia system. Use the solution specified
in the instructions when cleaning.
•
Mild detergent
•
70% isopropyl alcohol
•
Water
Compatible Disinfecting Solutions
The disinfecting solutions listed below are compatible for use on the Spectra Optia system.
•
0.63% sodium hypochlorite
•
70% isopropyl alcohol
•
0.50% ammonium chloride
If the disinfecting solution used by your facility contains an active ingredient that is not on this list, test the
solution before use by applying it to an inconspicuous area on the system to confirm that is does not damage or
discolor the system.
Warning:
Turn off the system before cleaning or disinfecting to prevent possible electrical shock or damage to
the equipment.
Warning:
When cleaning and disinfecting equipment surfaces that might have been exposed to blood, take
adequate precautions to prevent possible exposure to and transmission of infectious diseases.
Caution:
Clean the Spectra Optia system using a mild, non-abrasive cleaning solution or a mild detergent.
Industry standard practice defines mild cleaning solution and mild detergent as a solution or detergent that is safe
on skin and on washable surfaces, such as dishwashing liquid. Use of a solution that is not compatible with the
materials on the system may damage the material or operating characteristics of the system.
Caution:
Disinfect the Spectra Optia system using a disinfecting solution that is compatible for use on the
system. Use of a disinfecting solution that is not compatible could damage the material or operating characteristics
of the system.
SpectraOptiaOpsManual.book Page 230 Thursday, May 17, 2018 11:24 AM