Continuous Mononuclear Cell Collection (CMNC) Procedures
150
Spectra Optia Apheresis System Operator’s Manual
Entering and Changing Data During the Run
The Spectra Optia system requires that you enter specific data, such as patient data, to perform a CMNC
procedure. The system also allows you to change numerical values for certain operating parameters during the run.
Table 11-2 shows data entry ranges for the operating parameters according to the screen on which each data entry
button appears. To change or enter data, touch the button for the parameter that you want to change. A data entry
pad appears to enable you to enter a number. The data entry pad displays either the established range for the
parameter or the calculated range, if the range changes based on related data. Also, the range for Caution status is
displayed in yellow, if the data you enter could cause the system to operate in Caution status.
Table 11-2:
Data entry ranges for CMNC procedure parameters
Screen
Parameter
Range
Patient data
Sex
• Male
• Female
Height:
• ft
• in
• cm
• 1 to 8
• 0 to 11
• 30 to 244
Weight:
• lb
• kg
• 5 to 500
• 2 to 227
TBV (mL)
300 to 15000
Hct (%)
10 to 70
Options
Rinseback
• Yes
• No
Default:
• Yes if a custom prime is not
selected
• No if a custom prime is selected
Custom prime
• Yes
• No
Default: No
Saline rinse
• Yes
• No
Default: No
Blood warmer: Return line
• Yes
• No
Blood warmer: Tubing set volume (mL)
1 to 100
Plasma collection
• Now
• End of Run
Custom prime
Fluid type
• RBC
• Plasma
• Albumin
RBC unit Hct (%)
25 to 80
Inlet flow rate (mL/min)
5 to 70