![Terumo BCT Spectra Optia Apheresis System Скачать руководство пользователя страница 68](http://html1.mh-extra.com/html/terumo-bct/spectra-optia-apheresis-system/spectra-optia-apheresis-system_operators-manual_1090434068.webp)
Configuring the Spectra Optia System
54
Spectra Optia Apheresis System Operator’s Manual
Mononuclear Cell (MNC) Collection
Tables 4-8 and 4-9 show the operating parameters to configure for an MNC collection procedure.
Table 4-8:
Parameters and setting options to configure for an MNC collection procedure
Tab
Parameter
Description
Setting Options
MNC
Run target: Whole blood
processed (mL)
Amount of the patient’s blood to be processed. Does not
include AC.
If you enter a value for this parameter, the system
automatically sets the targets for the run time and the
TBV processed to zero.
100 to 50000
Default: 0
Run target: Run time (min)
Time required to complete the run.
The run starts when the patient is connected and ends
when the run targets have been attained. When you
enter a value for this parameter, the system
automatically sets the targets for the whole blood
processed and the TBV processed to zero.
40 to 480
Default: 0
Run target: TBV processed
Number of the patient’s total blood volumes to be pro-
cessed.
If you enter a value for this parameter, the system
automatically sets the targets for the whole blood
processed and the run time to zero.
1 to 10
Default: 2
Plasma: Plasma bag (mL)
Volume of plasma to be collected into the plasma bag.
0 to 500
Default: 0
Plasma: Collection
bag (mL)
Volume of plasma to be collected into the collection bag. 0 to 500
Default: 0
Chamber flush (mL)
Volume of plasma the system uses to flush cells from
the chamber into the collection bag during the collection
phase.
6 to 100
Default: 16
Chamber chase (mL)
Volume of plasma the system uses to rinse any cells
remaining in the collect line between the chamber and
the collect valve into the collection bag.
The chamber chase occurs after the chamber flush and
before the collect valve moves into the return position.
0 to 100
Default: 2
SpectraOptiaOpsManual.book Page 54 Thursday, May 17, 2018 11:24 AM