Spectra Optia Apheresis System Operator’s Manual
183
White Blood Cell Depletion (WBCD) and Platelet Depletion (PLTD) Procedures
13
Main run
Packing factor: WBCD procedures
• HES: Yes
• HES: No
1 to 20
• Default: 1.6
• Default: 4.5
Packing factor: PLTD procedures
1 to 20
Default: 20
Inlet flow rate (mL/min)
5 to 142
Collect flow rate (mL/min)
0.5 to 25
Collect volume (mL)
5 to the patient’s plasma volume
AC infusion rate (mL/min/L TBV)
0.2 to 2.5
(Caution status applies if the entry
exceeds 1.2.)
Inlet:AC ratio (__:1)
2 to 50
(Warning appears if the entered ratio
is not within the recommended range
of 6:1 to 15:1.)
Collection status
Collection preference
10 to 90
Default: 75 after the interface is
established
Run targets
Run time (min)
40 to 480
Whole blood processed (mL)
Patient’s TBV multiplied by 0.5 to 5
TBV processed
0.5 to 5
Collection bag (mL)
5 to the patient’s plasma volume
Rinseback
Return flow rate (mL/min)
2 to 100
Default: inlet flow rate that was in
use when the run ended or
100 mL/min, whichever is slower
Table 13-2:
Data entry ranges for WBCD and PLTD procedures (continued)
Screen
Parameter
Range