Spectra Optia Apheresis System Operator’s Manual
251
Spectra Optia System Specifications
Reservoir volume
• Between the high- and low-level sensors:
49 mL
• Entire reservoir: 56 mL
Reservoir prevents more than 0.5 mL of air
from entering the return line.
Sterile barrier filter
Collection Set and IDL Set:
• 0.2 microns.
• Helps to maintain a functionally closed
system.
Filter is on the AC line and on the saline line.
Reservoir filter
Prevents particles (200 microns or larger) from
entering the return line exiting the reservoir.
Inlet line trap
Traps large particles.
AC check valve
Prevents the free flow of AC through the tubing
set.
Administration line filter on BMP
Accessory Set
200 microns. Can be used for additional
filtration of bone marrow.
Filter is used for additional filtration of bone
marrow, if desired.
RBC residual volume
• When rinseback is performed (average
volume):
• Exchange Set: 10 mL
• Collection Set: 10 mL
• IDL Set: 17 mL
• When rinseback is not performed: Equal to
the ECV of the tubing set multiplied by the
patient’s Hct
Normal operating conditions
Channel life
• Exchange Set: 6 hours
• Collection Set: 8 hours
• IDL Set: 8 hours
Normal operating conditions
Table 18-26:
Tubing set and accessory set components (continued)
Characteristics
Performance
Conditions