Spectra Optia System Specifications
242
Spectra Optia Apheresis System Operator’s Manual
Spectra Optia System Specifications
General System Specifications
Electrical
Environmental
Table 18-1:
Electrical power and safety
Characteristics
Performance
Conditions
Electrical power rating
100 V AC to 240 V AC
50/60 Hz, 1050 VA
Where a circuit breaker provides excess
current protection
Ethernet port
Port is electrically isolated.
• Location where a device can be connected
to the system to print reports or to collect
and transmit data to Terumo BCT
• Location where service personnel can
connect to the system to download
procedure information to a computer when a
patient is not connected
Safety certifications
Meets the requirements of EN 60601-1 and
EN 61010-2-20
CE Marking
Device is certified by CSA International in
accordance with applicable U.S. and Canadian
standards. For applicable standards for the
product, see Table 18-36 on page 261.
CSA Certification to applicable U.S. and
Canadian standards
Meets the requirements of category IP21 of
IEC 60529
Table 18-2:
Environmental: Spectra Optia device
Characteristics
Performance
Conditions
Ambient operating
temperature
15.5 °C to 27.7 °C (60 °F to 82 °F)
Ambient operating
humidity
8% to 80%
Relative humidity (RH), non-condensing
Cleaning
• Surface of the device can be cleaned and
disinfected
• Pump rotors can be removed for cleaning
and disinfection
• Channel leaks are contained within the
centrifuge chamber
• Centrifuge chamber can be cleaned and
disinfected
Device is not damaged by cleaning and
disinfecting with the compatible cleaning
solutions and disinfecting solutions, according
to the instructions described in Chapter 17 of
this manual.
Access by foreign material
Device is protected against access by foreign
objects and falling water.
SpectraOptiaOpsManual.book Page 242 Thursday, May 17, 2018 11:24 AM