8.7 Message List
158
:COMParator:FLIMit:COUNt?
Query of upper limit and lower limit values of
first parameter for comparator function in
count value mode
*1, 2
:COMParator:FLIMit
:DEViation
<Reference Value>,
<Lower Limit Value>,
<Upper Limit Value>
<Reference Value>
= Numeric Value from
-199999 to 999999
(excluding 0)
(NR1)
<Lower Limit Value>,
<Upper Limit Value>
= OFF/Numeric Value from
-999.99 to 999.99 (NR2)
Setting of reference value and upper limit and
lower limit values of first parameter for com-
parator function in deviation percent mode
*2, 3
:COMParator:FLIMit
:DEViation?
Query of reference value and upper limit and
lower limit values of first parameter for com-
parator function in deviation percent mode
*1, 2
:COMParator:SLIMit:COUNt
<Lower Limit Value>,
<Upper Limit Value>
OFF/Numeric Value from
-199999 to 199999
(NR1)
Setting of upper limit and lower limit values of
second parameter for comparator function in
count value mode
*2, 3
:COMParator:SLIMit:COUNt?
Query of upper limit and lower limit values of
second parameter for comparator function in
count value mode
*1, 2
:COMParator:SLIMit
:DEViation
<Reference Value>,
<Lower Limit Value>,
<Upper Limit Value>
<Reference Value>
= Numeric Value from
-199999 to 199999 (NR1)
<Lower Limit Value>,
<Upper Limit Value>
= OFF/Numeric Value from
-199999 to 199999 (NR1)
Setting of reference value and upper limit and
lower limit values of second parameter for
comparator function in deviation percent
mode
*2, 3
:COMParator:SLIMit
:DEViation?
Query of reference value and upper limit and
lower limit values of second parameter for
comparator function in deviation percent
mode
*1, 2
Open Circuit and Short Circuit Compensation
:CORRection:DATA?
Query of compensation values for open circuit
and short circuit compensation
*1, 2
:CORRection:OPEN
ALL/ ON/ OFF/ RETurn
Setting of open circuit compensation function *2, 3
Unique Commands
Command
Data Section
Explanation
Error
Refer-
ence
Page
Error Explanations (An error is generated when a message is executed in the following cases)
*1 Query Error _______ When a response message exceeds 10 kB
*2 Execution Error ____ When a command is executed while open circuit, short circuit, or load compensation
*3 Execution Error ____ When set to other than the specified character data or numeric data.
*4 Execution Error ____When a number that has not been saved is specified.
*5 Execution Error ____When a command is executed during comparator measurement or BIN measure-
ment.
*6 Execution Error ____When not even one measurement value is saved to memory.
*7 Execution Error ____When there is an RS-232C specific command or query while the interface type is set
to GP-IB.
*8 Execution Error ____When a command to display a reference value in the SUB display area is executed
while the count setting is configured.
Note: Command errors are generated for all messages with a misspelling.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......