8.9 Message Reference
233
8
Setting and Query of Measurement Value Memory Function
ON/ OFF
Syntax
Command
Query
Response
:MEMory:CONTrol
<ON/ IN/ OFF>
:MEMory:CONTrol?
<ON/ IN/ OFF>
ON
: Save measurement values in memory.
IN
: Only measurement values whose parameters are evaluated
as good in the comparator and BIN function will be saved in
memory. (value will not be saved even if 1 comparator result
is Hi or Lo, or when the BIN result is OUT-OF-BINS or D-NG).
Saves measurement values in memory under normal mea-
surements.
OFF
: Measurement values not saved in memory.
Explanation
Command Sets the measurement value memory function.
When the setting is changed all saved measurement values are
erased.
Query
Returns the setting of the measurement value memory function as
characters.
Example
Command
Query
Response
:MEMory:CONTrol ON
Sets so that the measurement value is saved each time a measure-
ment is made.
:MEMory:CONTrol?
:MEMORY:CONTROL ON
(when HEADER ON)
ON
(when HEADER OFF)
The device is set to save the measurement value each time a mea-
surement is made.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......