5.1 Comparator Function
70
12.
Enter the D reference value using the arrow keys or the numeric
keypad.
Settable Range: -199999 to 199999
Move to the digit ........
and
Change the number ..
and
The reference value is set using the count value. At time of shipment it is set to ”0”.
13.
Press
to confirm the D reference value.
The reference value entered this time is not confirmed unless
is
pressed. The reference value used last time becomes valid.
14.
Press
to enter the D upper limit value setting mode.
15.
Enter the D upper limit value using the arrow keys or the numeric
keypad.
Settable Range: OFF,
-199999 to 199999 (When setting to "OFF"
)
Move to the digit ........
and
Change the number ..
and
Setting a value lower than the lower limit value will not produce an error, but
accurate judgment will not be possible.
At time of shipment, it is set to ” - - - - - - ” (OFF).
Light up
Flash
Upper limit value
(SUB display area )
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......