4.1 Open Circuit Compensation and Short Circuit Compensation
39
4
Take the open circuit compensation and
short circuit compensation values ________________________________
1.
In normal measurement mode, press
.
2.
Open the space between the HIGH and LOW terminals of the probe or fixture
connected to the measurement terminals to match the width of the object
being measured. (Connect Hc and Hp. Also connect Lc and Lp.)
3.
Use
or
to select setting the open circuit menu item.
Pressing
or
causes the display to change in the following manner.
(Flash)
(MAIN display area)
Flash
If you do not want to perform open circuit compensation, press
to pro-
ceed to configuring the short circuit compensation settings. (The SHORT LED
flashes and the “Short AdJuSt” indication is displayed and the “AdJuSt” indica-
tion flashes in the MAIN display area.) Proceed to Step 5.
(Example)
“
AdJuSt”
Delete invalid open adjustment values, after acquiring open
adjustment values only with measurement conditions set at
:CORRection:OPEN:POINt (Initial value: all measurement
conditions), the open adjustment values will be enabled.
See
"Setting and Query of Open Compensation Points" (p. 201)
“on”
Enable all disabled open adjustment values.
“oFF”
Disable open adjustment values at enabled measurement
values (frequency, level)
• When performing compensation, the placement of things like the probe and
the distances between terminals must be as similar as possible to the state
when performing measurement.
• If compensation is being affected by external noise, use the shielding pro-
cess.
See
Appendix 2 "Measurement of High Impedance Components" (p. A3)
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......