3.3 Setting the Measurement Conditions
29
9
3
You may set an equivalent circuit mode (SER/ PAR).
Automatic selection is also possible.
See
3.3.5 "Equivalent Circuit Mode" (p. 29)
Press
to change the mode.
Equivalent circuit mode:
AUTO
,
SER
,
PAR
The selected item is indicated by the lit LED lamp.
3.3.5 Equivalent Circuit Mode
AUTO
The series equivalent circuit mode or parallel equivalent circuit
mode is automatically selected according to the measurement
range.
SER
Series equivalent circuit mode
PAR
Parallel equivalent circuit mode
Range No.
Automatically selected mode
6 to 10
Series equivalent circuit
1 to 5
Parallel equivalent circuit