3.3 Setting the Measurement Conditions
26
Select a measurement mode.
Press
to change the mode.
Mode:
NORM
,
COMP
,
BIN
The selected item is indicated by the lit LED lamp.
Set the measurement frequency.
Set a frequency appropriate for the sample to be measured.
Press
to change the mode.
Measurement frequency:
120 Hz
,
1 kHz
The selected item is indicated by the lit LED lamp.
3.3 Setting the Measurement Conditions
3.3.1 Measurement Mode
NORM
Select this when using normal measurement mode.
COMP
Select this when using comparator measurement mode.
See
5.1 "Comparator Function" (p. 61)
BIN
(Model 3504-50,
3504-60 only)
Select this when using BIN measurement mode.
See
5.2 "BIN Measurement Function (Model 3504-50, 3504-60
The measurement conditions cannot be changed in comparator mode and
BIN mode. Set them in normal measurement mode.
3.3.2 Measurement Frequency
Содержание 3504-40 C HiTester