8.9 Message Reference
203
8
Setting and Query of Short Compensation Values
Syntax
Command
Query
Response
:CORRection:SHORt:DATA
<Compensation values1>
,
<Com-
pensation values 2>
:CORRection:SHORt:DATA?
<Compensation values 1>
,
<Compensation values 2>
<Compensation values 1>
= -99.9999E9 to 99.9999E9 (NR3)
<Compensation values 2>
= -99.9999E9 to 99.9999E9 (NR3)
Explanation
Command Sets the short compensation value for current measurement settings
(frequency, level).
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
Query
Returns as a numerical value the short compensation value for cur-
rent measurement settings (frequency, level).
Note
• The forwarding format for making settings and reading data with this
command follows the :CORRection:SHORT:DATA:FORMat setting.
When the forwarding format is ZPH
<Compensation values 1>
: Z
(Z
0 an execution error)
<Compensation values 2>
: PH
(PH
-180, PH
180 an execution error)
When the forwarding format is RSX
<Compensation values 1>
: Rs
<Compensation values 2>
: X
When the forwarding format is LSRS
<Compensation values 1>
: Ls
<Compensation values 2>
: Rs
• When the compensation value is-1E-21 to 1E-21 the setting value
will be 0.
Example
Command
:CORRection:SHORt:DATA:FORMat RSX
:CORRection:SHORt:DATA 5.67891E-03,3.34564E-05
When the forwarding format setting is RSX, the open compensation
value for the current measurement conditions will be set to
5.67891E-03,3.34564E-05
(Rs,X).
Query
:CORRection:SHORt:DATA?
:CORRection:SHORt:DATA 5.67891E-03,3.34564E-05
(when HEADER ON)
5.67891E-03,3.34564E-05
(when HEADER OFF)
When the forwarding format setting is RSX, the open compensation
value for the current measurement conditions is set to
5.67891E
-03,3.34564E-05
(Rs,X).
Note
• If the value forwarded to this device when converted to R and X is
outside of the -99.9999E9 to 99.9999E9 range an execution error
will occur.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......